Alternative method of ultrafiltration after cardiopulmonary bypass.
Document Type
Article
Publication Date
8-1-1994
Journal Title
The Annals of thoracic surgery
MeSH Headings
Adult, Blood, Cardiopulmonary Bypass, Child, Heart Arrest, Induced, Hemofiltration, Humans
Abstract
A revised circuit design for modified ultrafiltration is presented rendering the technique more convenient for use after cardiopulmonary bypass when blood cardioplegia is used. The procedure employs a hollow-fiber ultrafiltration device attached to the cardioplegia circuit. A bubble trap, heat exchanger, and a pressure monitor are incorporated as safety features. The technique has been used in 80 patients (30 pediatric and 50 adult) and has been associated with relevant increases in colloid osmotic pressure and hematocrit.
ISSN
0003-4975
First Page
573
Last Page
574
Recommended Citation
Groom, R C; Akl, B F; Albus, R A; Hill, A; Munoz, R; and Lefrak, E A, "Alternative method of ultrafiltration after cardiopulmonary bypass." (1994). MaineHealth Maine Medical Center. 864.